Chapter 2 Command System
:BUS<n>:POSition
Syntax
:BUS<n>:POSition <pos>
:BUS<n>:POSition?
Description
Sets or queries the vertical position of the bus on the screen.
Parameter
Name
<n>
<pos>
Return
The query returns an integer ranging from -167 to +217.
Format
Example
:BUS1:POSition 200
:BUS1:POSition?
:BUS<n>:THReshold
Syntax
:BUS<n>:THReshold <value>,<type>
:BUS<n>:THReshold? <type>
Description
Sets or queries the threshold of the specified decoding source.
Parameter
Name
<n>
<value>
<type>
Remarks
For VerticalScale, refer to the
the
For RS232 decoding, only when the Rx source is enabled, can you set the threshold.
For SPI decoding, only when the MOSI source is enabled, can you set the threshold.
Return
The query returns the threshold of the specified decoding source in scientific notation.
Format
Example
:BUS1:THReshold 2.4,PAL
:BUS1:THReshold? PAL
MSO7000/DS7000 Programming Guide
Type
Range
Discrete
{1|2|3|4}
Integer
-167 to +217
/*Sets the vertical position of the bus to 200.*/
/*The query returns 200.*/
Type
Range
Discrete
{1|2|3|4}
(-5× VerticalScale–OFFSet) to
Real
(5× VerticalScale–OFFSet)
{PAL|TX|RX|SCL|SDA|CS|CLK|MISO|MO
Discrete
SI|LIN|CAN|CANSub1}
:CHANnel<n>:OFFSet
command.
/*Sets the threshold of the PAL decoding source to 2.4 V.*/
/*The query returns 2.400000E0.*/
:CHANnel<n>:SCALe
command. For OFFSet, refer to
RIGOL
Default
——
Related to the
channel offset
Default
——
0
——