Math<N>:Offset; Math<N>:Invert; Math<N>:Reset - Rigol DS7034 User Programming Manual

Mso7000 series; ds7000 series; digital oscilloscope
Hide thumbs Also See for DS7034:
Table of Contents

Advertisement

Chapter 2 Command System
:MATH<n>:OFFSet
Syntax
:MATH<n>:OFFSet <offset>
:MATH<n>:OFFSet?
Description
Sets or queries the vertical offset of the operation results. The unit is related to the
currently selected operator and the unit selected by the source.
Parameter
Name
<n>
<offset>
Return
The query returns the vertical offset of the current operation results in scientific
Format
notation.
Example
:MATH1:OFFSet 8
:MATH1:OFFSet?
:MATH<n>:INVert
Syntax
:MATH<n>:INVert <bool>
:MATH<n>:INVert?
Description
Enables or disables the inverted display of the operation results; or queries the status
of the inverted display of the operation results.
Parameter
Name
<n>
<bool>
Remarks
This command is invalid for FFT operation.
Return
The query returns 1 or 0.
Format
Example
:MATH1:INVert ON
:MATH1:INVert?
:MATH<n>:RESet
Syntax
:MATH<n>:RESet
Description
After you send this command, the instrument will adjust the vertical scale of the
operation results to an optimal value based on the currently selected operator and the
horizontal time base of the source.
MSO7000/DS7000 Programming Guide
Type
Range
Discrete
{1|2|3|4}
Real
-1 GV to +1 GV
/*Sets the vertical offset to 8 V.*/
/*The query returns 8.000000E0.*/
Type
Range
Discrete
{1|2|3|4}
Bool
{{1|ON}|{0|OFF}}
/*Enables the inverted display.*/
/*The query returns 1.*/
RIGOL
Default
——
0.00 V
Default
——
0|OFF
2-97