Display:grading:time; Display:wbrightness; Display:grid - Rigol DS7034 User Programming Manual

Mso7000 series; ds7000 series; digital oscilloscope
Hide thumbs Also See for DS7034:
Table of Contents

Advertisement

RIGOL

:DISPlay:GRADing:TIME

Syntax
:DISPlay:GRADing:TIME <time>
:DISPlay:GRADing:TIME?
Description
Sets or queries the persistence time. The default unit is s.
Parameter
Name
<time>
Remarks
MIN: sets the persistence time to its minimum value to view how the waveform
changes at a high refresh rate.
specified value (e.g. 0.1, 0.2, 0.5, 1, 2, 5, 10): sets the persistence time to any of
the above specific value to observe glitch that changes relatively slowly or glitch
with low occurrence probability.
INFinite: In this mode, the oscilloscope displays the waveform newly acquired
without clearing the waveforms acquired formerly. It can be used to measure
noise and jitter and to capture incidental events.
Return
The query returns MIN, 0.1, 0.2, 0.5, 1, 2, 5, 10, or INF.
Format
Example
:DISPlay:GRADing:TIME 0.1
:DISPlay:GRADing:TIME?

:DISPlay:WBRightness

Syntax
:DISPlay:WBRightness <time>
:DISPlay:WBRightness?
Description
Sets or queries the brightness of the waveforms on the screen, expressed in
percentage.
Parameter
Name
<time>
Return
The query returns an integer ranging from 1 to 100.
Format
Example
:DISPlay:WBRightness 50
:DISPlay:WBRightness?

:DISPlay:GRID

Syntax
:DISPlay:GRID <grid>
:DISPlay:GRID?
Description
Sets or queries the display type of the screen grid.
Parameter
Name
<grid>
Remarks
FULL: turns on the background grid and coordinates.
HALF: turns the background grid off and turns the coordinate on.
NONE: turns the background grid and coordinate off.
IRE: the IRE command is only valid when the trigger type is Video trigger and the scale
2-66
Type
Range
Discrete
{MIN|0.1|0.2|0.5|1|2|5|10|INFinite}
/*Sets the persistence time to 0.1 s.*/
/*The query returns 0.1.*/
Type
Range
Integer
1 to 100
/*Sets the waveform brightness to 50%.*/
/*The query returns 50.*/
Type
Range
Discrete
{FULL|HALF|NONE|IRE}
Chapter 2 Command System
Default
MIN
Default
60
Default
FULL
MSO7000/DS7000 Programming Guide

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds7054Ds7024Ds7014Mso7034Mso7024Mso7014 ... Show all

Table of Contents