:MASK Commands

The :MASK commands are used to set or query the relevant parameters of the pass/fail test.
Command List:

:MASK:ENABle

:MASK:SOURce

:MASK:OPERate
:MASK:MDISplay
:MASK:X
:MASK:Y
:MASK:CREate
:MASK:RESet
:MASK:ENABle
Syntax
:MASK:ENABle <bool>
:MASK:ENABle?
Description
Enables or disables the pass/fail test function; or queries the on/off status of the pass/fail
test function.
Parameter
Name
<bool>
Remarks
The pass/fail test function is invalid in the following conditions: when the horizontal time
base mode is XY or ROLL; when in the slow sweep mode (i.g. the horizontal time base
mode is YT, and the horizontal time base is 200 ms/div or slower); when in waveform
recoding.
Return
The query returns 1 or 0.
Format
Example
:MASK:ENABle ON
:MASK:ENABle?
:MASK:SOURce
Syntax
:MASK:SOURce <source>
:MASK:SOURce?
Description
Sets or queries the measurement source of the pass/fail test.
Parameter
Name
<source>
Remarks
This command can only set the enabled channel, and you can send
the
:CHANnel<n>:DISPlay
Return
The query returns CHAN1, CHAN2, CHAN3, or CHAN4.
Format
Example
:MASK:SOURce CHANnel2 /*Sets the measurement source of the pass/fail test to CH2.*/
:MASK:SOURce?
2-90
Type
Range
Bool
{{1|ON}|{0|OFF}}
/*Enables the pass/fail test function.*/
/*The query returns 1.*/
Type
Range
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
command to enable the desired channel.
/*The query returns CHAN2.*/
Chapter 2 Command System
MSO7000/DS7000 Programming Guide
Default
0|OFF
Default
CHANnel1