RIGOL
Remarks
POSitive: indicates the rising edge.
NEGative: indicates the falling edge.
RFALl: indicates the rising or falling edge.
Return
The query returns POS, NEG, or RFAL.
Format
Example
:TRIGger:EDGE:SLOPe NEGative
:TRIGger:EDGE:SLOPe?
:TRIGger:EDGE:LEVel
Syntax
:TRIGger:EDGE:LEVel <level>
:TRIGger:EDGE:LEVel?
Description
Sets or queries the trigger level of Edge trigger. The unit is the same as that of current
amplitude of the selected source.
Parameter
Name
<level>
Remarks
For VerticalScale, refer to the
to the
Only when the selected source is an analog channel, a digital channel, or an
external trigger, can this setting command is valid.
Return
The query returns the trigger level in scientific notation.
Format
Example
:TRIGger:EDGE:LEVel 0.16
:TRIGger:EDGE:LEVel?
2-166
Type
Range
(-5 × VerticalScale - OFFSet) to
Real
(5 × VerticalScale - OFFSet)
:CHANnel<n>:OFFSet
/*Sets the trigger level to 160 mV.*/
/*The query returns 1.600000E-1.*/
/*Sets the edge type to Falling Edge.*/
/*The query returns NEG.*/
:CHANnel<n>:SCALe
command. For OFFSet, refer
command.
MSO7000/DS7000 Programming Guide
Chapter 2 Command System
Default
0