Rigol DS7034 User Programming Manual page 218

Mso7000 series; ds7000 series; digital oscilloscope
Hide thumbs Also See for DS7034:
Table of Contents

Advertisement

RIGOL
:TRIGger:IIC:CLEVel
Syntax
:TRIGger:IIC:CLEVel <level>
:TRIGger:IIC:CLEVel?
Description
Sets or queries the trigger level of the clock line in I2C trigger. The unit is the same
as that of the current amplitude.
Parameter
Name
<level>
Remarks
For VerticalScale, refer to the
the
:CHANnel<n>:OFFSet
Return
The query returns the trigger level in scientific notation.
Format
Example
:TRIGger:IIC:CLEVel 0.16
:TRIGger:IIC:CLEVel?
:TRIGger:IIC:DLEVel
Syntax
:TRIGger:IIC:DLEVel <level>
:TRIGger:IIC:DLEVel?
Description
Sets or queries the trigger level of the data line in I2C trigger. The unit is the same
as that of the current amplitude.
Parameter
Name
<level>
Remarks
For VerticalScale, refer to the
the
:CHANnel<n>:OFFSet
Return
The query returns the trigger level in scientific notation.
Format
Example
:TRIGger:IIC:DLEVel 0.16
:TRIGger:IIC:DLEVel?
2-204
Type
Range
(-5 × VerticalScale - OFFSet) to
Real
(5 × VerticalScale - OFFSet)
:CHANnel<n>:SCALe
command.
/*Sets the trigger level to 160 mV.*/
/*The query returns 1.600000E-1.*/
Type
Range
(-5 × VerticalScale - OFFSet) to
Real
(5 × VerticalScale - OFFSet)
:CHANnel<n>:SCALe
command.
/*Sets the trigger level to 160 mV.*/
/*The query returns 1.600000E-1.*/
Chapter 2 Command System
Default
0 V
command; for OFFSet, refer to
Default
0 V
command; for OFFSet, refer to