Rigol DS7034 User Programming Manual page 234

Mso7000 series; ds7000 series; digital oscilloscope
Hide thumbs Also See for DS7034:
Table of Contents

Advertisement

RIGOL
:TRIGger:LIN:BAUD
Syntax
:TRIGger:LIN:BAUD <baud>
:TRIGger:LIN:BAUD?
Description
Sets or queries the baud rate of LIN trigger. The default unit is bps.
Parameter
Name
<baud>
Remarks
If the baud rate is set to a value with "M", then "A" should be added at the end of the
value. For example, if you send 5 M, you need to send 5 MA.
The query returns an integer ranging from 1 kbps to 20 Mbps.
Return
Format
Example
:TRIGger:LIN:BAUD 19200
:TRIGger:LIN:BAUD?
:TRIGger:LIN:STANdard
Syntax
:TRIGger:LIN:STANdard <std>
:TRIGger:LIN:STANdard?
Description
Sets or queries the protocol version of LIN trigger.
Parameter
Name
<std>
Return Format
The query returns 1X or 2X.
Example
:TRIGger:LIN:STANdard 2X
:TRIGger:LIN:STANdard?
:TRIGger:LIN:SAMPlepoint
Syntax
:TRIGger:LIN:SAMPlepoint <value>
:TRIGger:LIN:SAMPlepoint?
Description
Sets or queries the sample position of LIN trigger.
Parameter
Name
<value>
Description
The sample position is expressed as the ratio of "time from the bit start to the sample
point" to "bit time", in %.
Return Format
The query returns an integer ranging from 10 to 90.
Example
:TRIGger:LIN:SAMPlepoint
:TRIGger:LIN:SAMPlepoint? /*The query returns CHAN2.*/
2-220
Type
Range
Integer
1 kbps to 20 Mbps
/*Sets the baud rate of LIN trigger to 19200 bps.*/
/*The query returns 19200.*/
Type
Range
{1X|2X}
Discrete
/*Sets the protocol version of LIN trigger to 2X.*/
/*The query returns 2X.*/
Type
Range
Integer
10 to 90
/*Sets the trigger source to CH2.*/
Chapter 2 Command System
MSO7000/DS7000 Programming Guide
Default
9600 bps
Default
1X
Default
50

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds7054Ds7024Ds7014Mso7034Mso7024Mso7014 ... Show all

Table of Contents