Chapter 2 Command System
:TRIGger:NEDGe:IDLE
Syntax
:TRIGger:NEDGe:IDLE <time>
:TRIGger:NEDGe:IDLE?
Description
Sets or queries the idle time of the Nth Edge trigger. The default unit is s.
Parameter
Name
<time>
Return
The query returns the idle time in scientific notation.
Format
Example
:TRIGger:NEDGe:IDLE 0.002
:TRIGger:NEDGe:IDLE?
:TRIGger:NEDGe:EDGE
Syntax
:TRIGger:NEDGe:EDGE <edge>
:TRIGger:NEDGe:EDGE?
Description
Sets or queries the number of edges of the Nth Edge trigger.
Parameter
Name
<edge>
Return
The query returns an integer ranging from 1 to 65535.
Format
Example
:TRIGger:NEDGe:EDGE 20
:TRIGger:NEDGe:EDGE?
:TRIGger:NEDGe:LEVel
Syntax
:TRIGger:NEDGe:LEVel <level>
:TRIGger:NEDGe:LEVel?
Description
Sets or queries the trigger level of the Nth Edge trigger. The unit is the same as that
of current amplitude.
Parameter
Name
<level>
Remarks
For VerticalScale, refer to the
the
:CHANnel<n>:OFFSet
Return
The query returns the trigger level in scientific notation.
Format
Example
:TRIGger:NEDGe:LEVel 0.16
:TRIGger:NEDGe:LEVel?
MSO7000/DS7000 Programming Guide
Type
Range
Real
16 ns to 1 s
/*Sets the idle time to 2 ms.*/
/*The query returns 2.000000E-3.*/
Type
Range
Integer
1 to 65535
/*Sets the number of edges to 20.*/
/*The query returns 20.*/
Type
Range
(-5 × VerticalScale - OFFSet) to
Real
(5 × VerticalScale - OFFSet)
:CHANnel<n>:SCALe
command.
/*Sets the trigger level to 160 mV.*/
/*The query returns 1.600000E-1.*/
Default
1 μs
Default
2
Default
0 V
command; for OFFSet, refer to