Histogram:blimit; Histogram:llimit; Histogram:rlimit - Rigol DS7034 User Programming Manual

Mso7000 series; ds7000 series; digital oscilloscope
Hide thumbs Also See for DS7034:
Table of Contents

Advertisement

Chapter 2 Command System

:HISTogram:BLIMit

Syntax
:HISTogram:BLIMit <y>
:HISTogram:BLIMit?
Description
Sets or queries the histogram's bottom boundary limit.
Parameter
Name
<y>
Return
The query returns the histogram's bottom boundary limit in scientific notation.
Format
Example
:HISTogram:BLIMit -2
:HISTogram:BLIMit?

:HISTogram:LLIMit

Syntax
:HISTogram:LLIMit <x>
:HISTogram:LLIMit?
Description
Sets or queries the histogram's left boundary limit.
Parameter
Name
<x>
Return
The query returns the histogram's left boundary limit in scientific notation.
Format
Example
:HISTogram:LLIMit -2
:HISTogram:LLIMit?

:HISTogram:RLIMit

Syntax
:HISTogram:RLIMit <x>
:HISTogram:RLIMit?
Description
Sets or queries the histogram's right boundary limit.
Parameter
Name
<x>
Return
The query returns the histogram's right boundary limit in scientific notation.
Format
Example
:HISTogram:RLIMit -2
:HISTogram:RLIMit?
MSO7000/DS7000 Programming Guide
Type
Range
-4× VerticalScale to
4× VerticalScale
Real
(vertical scale of the channel
source of the histogram)
/*Sets the histogram's bottom boundary limit to -2 mV.*/
/*The query returns -2.000000E0.*/
Type
Range
(-5× Horizontal Time Base) to
Real
(5× Horizontal Time Base)
/*Sets the histogram's left boundary limit to -2 mV.*/
/*The query returns -2.000000E0.*/
Type
Range
(-5× Horizontal Scale) to
Real
(5× Horizontal Scale)
/*Sets the histogram's left boundary limit to -2 mV.*/
/*The query returns -2.000000E0.*/
Default
0 V
Default
0 V
Default
0 V
RIGOL
2-73

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds7054Ds7024Ds7014Mso7034Mso7024Mso7014 ... Show all

Table of Contents