RIGOL
:TRIGger:WINDows:POSition
Syntax
:TRIGger:WINDows:POSition <pos>
:TRIGger:WINDows:POSition?
Description
Sets or queries the trigger position of Window trigger.
Parameter
Name
<type>
Remarks
EXIT: triggers when the input signal exits the specified trigger level range.
ENTER: triggers when the input signal enters the specified trigger level range.
TIME: triggers when the accumulated hold time after the trigger signal enters the
specified trigger level range is equal to the window time.
Return
The query returns EXIT, ENTER, or TIM.
Format
Example
:TRIGger:WINDows:POSition ENTER
:TRIGger:WINDows:POSition?
:TRIGger:WINDows:TIME
Syntax
:TRIGger:WINDows:TIME <time>
:TRIGger:WINDows:TIME?
Description
Sets or queries the window time of Window trigger.
Parameter
Name
<time>
Return
The query returns the window time in scientific notation.
Format
Example
:TRIGger:WINDows:TIME 0.002
:TRIGger:WINDows:TIME?
:TRIGger:WINDows:ALEVel
Syntax
:TRIGger:WINDows:ALEVel <level>
:TRIGger:WINDows:ALEVel?
Description
Sets or queries the upper limit of the trigger level of Window trigger. The unit is the
same as that of the current amplitude.
Parameter
Name
<level>
Remarks
For VerticalScale, refer to the
the
:CHANnel<n>:OFFSet
Return
The query returns the upper limit of the trigger level in scientific notation.
Format
Example
:TRIGger:WINDows:ALEVel 0.16 /*Sets the upper limit of the trigger level to 160 mV.*/
:TRIGger:WINDows:ALEVel?
2-186
Type
Range
Discrete
{EXIT|ENTER|TIME}
Type
Range
Real
8 ns to 10 s
/*The query returns 2.000000E-3.*/
Type
Range
Real
Lower limit to (5 × VerticalScale - OFFSet)
:CHANnel<n>:SCALe
command.
/*The query returns 1.600000E-1.*/
/*Sets the trigger position to ENTER.*/
/*The query returns ENTER.*/
/*Sets the window time to 2 ms.*/
command; for OFFSet, refer to
MSO7000/DS7000 Programming Guide
Chapter 2 Command System
Default
ENTER
Default
1 μs
Default
0 V

Advertisement