Data Communications & Alarm Interface
All signal connections are installed in the electrical interface panel at the rear of the incubator.

RS-232 Interface

The RS- 232 interface (item 2 in
incubators to the serial interface port of a computer to allow for the computer-aided
acquisition and documentation of major operating parameters (temperature, error codes, etc.).
[1]
Sample Sensor connection (only IMH-S)
[2]
RS 232 interface
[3]

Alarm contact

[4]
Power socket
[5]
2 fuses 2 A, for socket and lamp

Figure 4-11 Signal Interfaces and Power Socket

Alarm Contact
The incubator can be connected to an on-site, external alarm system (such as a private branch
telephone exchange, a facility monitoring system, visual or audible alarm indicators). For this
purpose, the devices come with a pre-wired potential-free alarm contact (item 3 in
figure
4-11)
loop or the incubator's electrical circuits or hardware.

AC Power Socket

The incubator is connected to the AC supply mains via the socket (item 4 in
which accepts a power cord with an IEC standard plug.

Fuses

Two 16 A slow-blow fuses mounted on the incubator's main electronic circuit board protect
internal circuitry from the impact of excessive power consumption.
Thermo Scientific
figure 4-11
The alarm contact is energized whenever an error occurs in an internal control
.
Data Communications & Alarm Interface
below) may be used to connect Heratherm
Product Description
figure
4-11),
Heratherm
4-13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents