13.Waveform Comparison / Waveform and Displacement Comparison Mode

13-7-2. Setting Method

1. Set "Channel Number Selection" in "Code Setting" to the same operation ch as used
for measurement.
2. Set the HI-HI Limit / LO-LO Limit judgment to "Hold". (If set at "Sampling ", real-
time Judgments on the sampling value will result.)
3. Set the hold to any of "Sample Hold", "Peak Hold" and "Valley Hold". If the hold is
set at "None", real-time High/Low Limit Judgments on the sampling value with the
HI-HI Limit / LO-LO Limit will result.
4. For High/Low Limit Judgment of the hold point, set the HI-HI Limit / LO-LO Limit.
(See "16-2-7. High Limit / Low Limit / HI-HI/LO-LO Mode / HI-HI Limit / LO-LO
Limit" P109.)

13-8. Un-passing Area Output

If the Un-passing Area Output setting (See "16-2-17. Un-passing Area Output" P120.) is
"OFF", the OK signal is not output when sampling does not pass the comparison area.
Since the judgment output is not made until sampling enters the comparison area in this
case, if it ends in an un-passing state, all of the HI, OK and LO outputs are turned off.
When the setting is "ON"
Always acceptable
High limit
Low limit
Start point End point
Comparison area
The results are maintained.
Input waveform
80