5-5. LOCK Release

Calibrated values and set values can be locked to prevent a change due to misoperation.
There are two types of locks: a software lock to be operated from the screen, and a
hardware lock by the switch on the rear panel. For performing calibration, release both
locks.
1. Turn off the LOCK switch on the rear panel. (Hardware lock)
2. Set the "Set Value Protection" setting to "All Unprotect". (Software lock)
Now, the locks are released. Upon completion of calibration, hardware lock and software
lock again to prevent misoperation.

5-6. Excitation Voltage

Set the voltage value to be applied to the bridge of the strain gauge sensor. Select as large
a value as possible within the range not exceeding the maximum excitation voltage of the
sensor.
Setting item: 2.5V / 10V
The excitation voltage of the F395 is 2.5V/10V. If the maximum
excitation voltage of the sensor is less than 2.5V/10V, heating or
breakage may result.
LOCK
OFF
MODE
System →
CH1
MODE
CAL.
CH2
Protect
All Unprotect
EXC. Volt
CAUTION
5.Methods of Calibration