With respect to the X-axis
Waveform Start Level
START
When the Level Axis Select setting is "X-axis"
・ The level edge of "Waveform Start Level" occurs only when the X-axis
measured value exceeds the set value in the direction of negative →
positive. Be aware that the level edge does not occur if positive →
negative. Also, the level=set value -1. (If 100 is set, sampling starts
from the X-axis=100 when the X-axis measured value operates as 98→
99 → 100).
・ If the Operation Mode is changed to the Multi-Hold / Waveform
Comparison mode, the setting is automatically changed to "Y-axis". (X-
axis cannot be set in the Multi-Hold / Waveform Comparison mode.)
・ When the Level Axis Select setting is changed from "Y-axis" to "X-
axis", if a negative value is set, you will hear a beep, and at the same
time, it is automatically changed to its default value.
・ The Waveform Termination Level is effective when the signal
"Waveform Termination Level Operation" is ON.
・ For validating the Waveform Termination Level in hysteresis return
measurement, it is recommended to input the hysteresis return signal
to the signal "Waveform termination level operation".
・ With respect to the X-axis, Digital Zero by a sampling start is not
performed on the displacement sensor side.
・ The displacement digital-zero function can operate only when at least
one lock is applied, such as CAL. Protect.
0
16.Code Setting
Waveform Termination Level
The solid line
shows a waveform
that is drawn
on the screen.
t
114