Unit; Increment; Decimal Place; Zero Calibration - Unipulse f395 Operation Manual

Dynamic force processor
Table of Contents

Advertisement

5.Methods of Calibration

5-7. Unit

Set the measuring unit. Select from the following.

5-8. Increment

Set the minimum value of digital changes.

5-9. Decimal Place

Set the decimal place.

5-10. Zero Calibration

Set the initial zero point.
1. Bring the sensor into a zero condition (no-load condition).
2. Press OK under Zero Calibration.
3. For error display, see "19. Error Messages" P150.
35
Setting item: kg / t / g / N / kN / MN / N・m / Pa / kPa / MPa / bar / N/m
/ μm / rad / deg / lb / oz / dyne / psig / ftlb / inlb / inoz / ft / in / None
CH1
MODE
CAL.
CH2
Setting range: 1 ~ 100 (Not including the decimal point.)
CH1
MODE
CAL.
CH2
Setting item: 0 / 0.0 / 0.00 / 0.000
CH1
MODE
CAL.
CH2
CH1
MODE
CAL.
CH2
PAGE
Unit
PAGE →
Increment
Decimal Place
PAGE
Decimal Place
Zero CAL.
2
/ mm / cm