5-3. Equivalent Input Calibration Procedures

Perform equivalent input calibration of the sensor as follows:
LOCK switch OFF
Set value protection cancel
Excitation voltage setting
Unit setting
Decimal place setting
Increment setting
Zero calibration
Equivalent input calibration
Set value protection
LOCK switch ON
Turn off the LOCK switch on the rear panel to unprotect
calibration.
Set the "Set Value Protection" setting to unprotect
calibration.
Select the "Excitation Voltage" from 2.5V/10V.
(Set 2.5V when the input specification is voltage / current.)
Turn off the power, and connect the sensor.
Set the "Unit" to be used.
Set the "Decimal Place".
Set the minimum value of the digital changes.
Set the initial zero point.
Set the rated output value of the sensor.
Protect calibration to prevent misoperation.
Turn on the LOCK switch to protect calibration to prevent
misoperation.