5-11. Equivalent Input Calibration

Set the rated output value of the sensor.
1. Select the "Equivalent Input Calibration" item.
2. Input set values as follows:
 ("i.- iii." depend on the analog input specifications (factory-shipped settings).)
i. Strain gauge type sensor
Sensor Output mV/V? .. Setting range: -9.999 ~ -0.100, 0.100 ~9.999
Load? ............................ Setting range: 100 ~ 9999 (Not including the decimal point.)
ii. Voltage output sensor
Sensor Output V?.......... Setting range: -9.999 ~ -0.100, 0.100 ~ 9.999
Load? ............................ Setting range: 100 ~ 9999 (Not including the decimal point.)
iii. Current output sensor
Sensor Output mA? ...... Setting range: -99.99 ~ -1.00, 1.00 ~ 99.99
Load? ............................ Setting range:100 ~ 9999 (Not including the decimal point.)
3. For error display, see "19. Error Messages" P150.
CH1
MODE
CAL.
CH2
The strain gauge type sensor comes with a data sheet upon purchase.
Capacity ................... Load/displacement (unit: kN, Mpa, etc.)
Rated Output............ Voltage (unit: mV/V)
Non-Linearity, Hysterisis,Input Resistance,Output Resistance,
Zero Balance, etc.
Values necessary for equivalent input calibration are the capacity and
rated output. Input these two values to the F395.
Equiv. CAL.
5.Methods of Calibration
36