5-11. Equivalent Input Calibration
Set the rated output value of the sensor.
1. Select the "Equivalent Input Calibration" item.
2. Input set values as follows:
("i.- iii." depend on the analog input specifications (factory-shipped settings).)
i. Strain gauge type sensor
Sensor Output mV/V? .. Setting range: -9.999 ~ -0.100, 0.100 ~9.999
Load? ............................ Setting range: 100 ~ 9999 (Not including the decimal point.)
ii. Voltage output sensor
Sensor Output V?.......... Setting range: -9.999 ~ -0.100, 0.100 ~ 9.999
Load? ............................ Setting range: 100 ~ 9999 (Not including the decimal point.)
iii. Current output sensor
Sensor Output mA? ...... Setting range: -99.99 ~ -1.00, 1.00 ~ 99.99
Load? ............................ Setting range:100 ~ 9999 (Not including the decimal point.)
3. For error display, see "19. Error Messages" P150.
CH1
MODE
CAL.
→
CH2
The strain gauge type sensor comes with a data sheet upon purchase.
Capacity ................... Load/displacement (unit: kN, Mpa, etc.)
Rated Output............ Voltage (unit: mV/V)
Non-Linearity, Hysterisis,Input Resistance,Output Resistance,
Zero Balance, etc.
Values necessary for equivalent input calibration are the capacity and
rated output. Input these two values to the F395.
Equiv. CAL.
→
5.Methods of Calibration
36