17.System-related Settings and Operations
17. System-related Settings and Operations

17-1. Set Value Protection

By this function, it is prohibited to change set values and calibrated values in order to
prevent them from being changed due to misoperation. The following three patterns can
be set.
All Unprotect .. Calibration can be performed, and also set values can be changed.
CAL. Protect ... Calibration cannot be performed, but set values can be changed.
All Protect ........ Calibration cannot be performed, and also set values cannot be changed.

17-2. Display Update Rate

Set the display update rate of measured values.

17-3. Backlight

The backlight of the color liquid crystal display (LCD) is automatically turned off by this
function. If the touch panel is not touched within the time (min.) set here, the backlight
of the LCD goes out. For lighting the backlight, touch the touch panel, or the turn on the
backlight ON signal.
(See "4-3-2. Input and Output Signals" P25.)
127
Setting item: All Unprotect / CAL. Protect / All Protect
MODE
System
The DZ (Digital Zero) function cannot be used.
(For the items applicable to "Calibration", see "24. Setting Items list"
P161.)
The DZ (Digital Zero) function can be used.
The DZ (Digital Zero) function can be used.
Setting range: 0.1 ~ 9.9
MODE
System
Setting range: 0 ~ 99
MODE
System
Protect
Display Rate
Backlight