16.Code Setting
The Hold Start Level is displayed only when Time Specified Automatic Hold (Peak, Val-
ley, P-P) / Relative Minimum Value Hold / Relative Maximum Value Hold / Inflection
Point Hold A ~ D is selected in Hold Setting to set the start level for executing each
hold. The setting is not displayed when any hold other than the above is used, or when
the Waveform Comparison / Waveform and Displacement Comparison / Hysteresis 2
mode is used.
Hold Start Level

16-2-10. Hold Detect Time

This is displayed only when Time Specified / Time Specified Automatic is selected in
Hold Setting to set the time of hold period.
The unit of Hold Detect Time is one-time sampling time. For example, if the hold detect
time is set to 1000 when the sampling rate in Operation Mode is set at 1kHz, 1 sec. will
result, and if it is set to 2500, 2.5 sec. will result. For details, see "7-2. Sampling Rate"
P40.
115
Setting range: -9999 ~ +9999
MODE → Code Set → Basic: Hold Set → Hold Start Level
If -9999 is set, the above hold operates with the Waveform Start Level.
(The Hold Start Level is ineffective.)
Setting range: 1 ~ 9999
MODE
Code Set → Basic: Hold Set → Hold Timer