Waveform Sampling Procedures - Unipulse f395 Operation Manual

Dynamic force processor
Table of Contents

Advertisement

13.Waveform Comparison / Waveform and Displacement Comparison Mode
2. Sample a waveform.
3. Edit the waveform as necessary.
4. Set the area.
5. Set the same value of Channel Number Selection set in "1." under Waveform Save.
6. By setting the channel having been set in "1." and "5." with channel selection

13-3. Waveform Sampling Procedures

Waveform sampling can be performed efficiently by the following procedures. (It is
assumed that the F395 has been calibrated.)
For the setting / operating method of each function, see "13-7. Hold Operation" P78.
(For details of input and output signals, see "4-3-2. Input and Output Signals" P25.)
* If no step is shown for each item, select the item by the step of "MODE → Waveform
1. Set "Wave Comp." or "Waveform&Displace" under "Operation Mode".
69
(Register the waveform.)
signals (CODE1 ~ 16: See "4-3-2. Input and Output Signals" P25.), measurements
can be carried out by using the values having been set in "1." and the waveform
registered in "5.". (Set the operation ch.)
・ If the operation ch is changed during measurement, the
measurement is stopped forcedly. (The results of previous
judgments are not displayed.)
・ For hold operation, input the T/H signal as necessary.
Sampling".
MODE → Operation Mode → Operation Mode → Wave Comp.
or
Wave&Displace