Unipulse f395 Operation Manual page 170

Dynamic force processor
Table of Contents

Advertisement

Voltage input
Signal input range:
Input impedance:
Zero/gain adjustment:
Equivalent input calibration range:
Equivalent input calibration range error:
Accuracy:
Analog filter:
A/D converter:
Current input
Signal input range:
Input resistance:
Zero/gain adjustment:
Equivalent input calibration range:
Equivalent input calibration range error:
Accuracy:
Analog filter:
-5V ~ +5V
5kΩ or more
By digital operation.
-5V ~ -1V, +1V ~ +5V
Within 0.2%F.S.
Non-linearity:
Within 0.02% F.S. ± 1 digit
(at 5V input).
Zero drift:
Within 50μV/ ℃ RTI
(input-converted value).
Gain drift:
Within 0.05%/ ℃ .
Selectable from 10, 50, 200, and 600Hz.
Speed:
4000 times/sec. (max.) when one input is used.
2000 times/sec. (max., X-axis measurement)
when two inputs are used.
Changeable to 100, 200, 500, 1000, 2000 times/sec.
according to the input waveform.
Resolution:
Approx. 1/27000 relative to 5V
-20mA ~ +20mA
Approx. 100Ω
By digital operation.
-20mA ~ -8mA
+8mA ~ +20mA
Within 0.2% F.S.
Non-linearity:
Within 0.02% F.S. ± 1 digit
(at 20mA input)
Zero drift:
Within 2μA/ ℃ RTI
(input-converted value).
Gain drift:
Within 0.05%/ ℃ .
Selectable from 10, 50, 200, and 600Hz.
23.Specifications