15-6. X-axis Start Point

Set the X-axis start point of the graph.
Setting range: Changed according to the Operation Mode, Sampling Rate, and Actual
X-axis
Start Point

15-7. X-axis Scaling

Set the X-axis scale of the graph. Select scaling so that the input waveform is properly
displayed.
Setting item: 100% / 50% / 20% / 10%
Load Calibration value for the X-axis.
(Not including the decimal point in the following table.)
(Load input value for Equivalent Input Calibration (pulse input
calibration for PUI))
Multi-Hold / Waveform
Comparison mode
Sampling rate
Range
4kHz
0 ~ 500
2kHz
0 ~ 1000
1kHz
0 ~ 2000
500Hz
0 ~ 4000
200Hz
0 ~ 10000
100Hz
0 ~ 20000
MODE
→ Read Wave →
MODE
Read Wave
Hysteresis / Hysteresis 2 / Waveform
and Displacement Comparison mode
Calibration value
1 ~ 2047
2048 ~ 4095
4096 ~ 8191
8192 ~ 16383
X StartPoint
X Scale
15.Waveform Reading
Range
0 ~ 2000
0 ~ 4000
0 ~ 8000
0 ~ 16000
X-axis scale
98