5.Methods of Calibration

5-12. Actual Load Calibration

Set the span (gain) point of the sensor.
1. Select the "Actual Load Calibration" item.
2. Set in accordance with the display.
3. For error display, see "19. Error Messages" P150.
37
CH1
MODE
CAL.
CH2
Load?.... Setting range: -9999 ~ -100, 100 ~ 9999 (Not including the decimal point.)
Calibration cannot be performed with the same input value
as zero calibration.
The full scale value of displacement is determined by the value
input for the "Load?" item of Equivalent Input Calibration or Actual
Load Calibration in the Hysteresis / Hysteresis 2 / Waveform and
Displacement Comparison mode. The relationship between the
input value (X-axis calibrated value) and full scale value is as
follows:
Input value
(X-axis calibrated value)
2048 ~ 4095
4096 ~ 8191
8192 ~ 9999
Actual CAL.
Full scale value
1 ~ 2047
2047
4096
8188
16376