Installation And Connections - Keithley S530 Administrative Manual

Parametric test system
Hide thumbs Also See for S530:
Table of Contents

Advertisement

S530 Administrative Guide
When you perform removal and installation procedures, or maintenance on the system, lockout by placing a
padlock through the bracket by the PDU breaker and tagout as appropriate (see the
for more information).
Remove system power
1. You must shut down the software, and remove all power from the computer and the system (see
down using ACS
WARNING
Before proceeding, you must make sure the power indicator on the front door is NOT
illuminated.
2. Place the power switch for the PDU to the OFF position. The location of the PDU is at the back of the
cabinet below the rear door.
3. If you are working in the system cabinet, disconnect the system cabinet line cord from the AC line power
receptacles.
WARNING
When you remove power, make sure that you disconnect all system cabinet line cords from
AC line power receptacles.
4. Verify that all power has been removed and discharged from the system cabinet by switching the main
power switch to the ON position (located on the front door of the cabinet) and verify that the green light
does not illuminate. If the light does not come on, the power is off. Turn main power switch back to the OFF
position.
5. Lockout and tagout the system source power connection in accordance with your company's lockout/tagout
policy.

Installation and connections

WARNING
The following installation and connection procedures should be performed by trained site
installers who are familiar with the associated physical and electrical hazards. Also, you
should never turn on the system until all connections and safety grounds are installed.
Position the system cabinet
The system cabinet contains the controller and instrumentation for the test system. The cabinet is on casters, which
allows you to easily roll on a hard floor surface. The two steering casters in the rear are swivel type, while the two
casters at the front are in fixed positions.
To position the system cabinets:
1. Carefully roll the system cabinet to its desired location next to the prober. Allow approximately 23.5 inches
(597 mm, nominal) of clearance between the cabinet and other instrumentation.
2. Lock the casters by pushing down on the caster-locking mechanisms located near the front-bottom of the
cabinet.
3. Adjust the height of the four legs so that the weight of the cabinet is on the legs and not on the casters.
Adjust the legs so that the cabinet is level and does not move (see the
PA-992 Rev. D / April 2012
or
Shut down using
KTE).
Remove system power
Seismic securement
topic).
topic
Shut
19