7.3.5
Glow Luminescence
Sensitivity
Measurement Parameters:
Plate Layout:
1
2
A
Bx
ATP
B
Bx
ATP
C
Bx
ATP
D
Bx
ATP
E
F
G
H
Material/Reagents:
2016-04
For the Infinite 200 PRO with the option 'Luminescence the following tests may
be performed to prove the sensitivity specifications:
Perform the following measurement to determine the sensitivity:
Parameter
Reading Mode
Integration Time
Settle Time
Plate Type
Part of the Plate
Pipette 200 µl of the ATP reagents into the appropriate wells according to the
plate layout:
3
4
5
Bx
B
B
Bx
B
B
Bx
B
B
Bx
B
B
-8
ATP
2*10
M ATP (final concentration in well)
B
Blank (ATP reagent: Tris-EDTA=1:5)
Bx
Blank (wells used for cross-talk calculation)
BioThema ATP Kit (ATP-Kit SL 144-041, BioThema AB)
1 Greiner 96-well plate white
200 µl Pipette + tips
IFU for Infinite 200 PRO No. 30052730
Setting
Luminescence
1000 ms
0
GRE96fw
A1 – D10
6
7
8
B
B
B
B
B
B
B
B
B
B
B
B
Rev. No. 1.6
7. Quality Control
9
10
11
B
B
B
B
B
B
B
B
12
169