Washing Procedure (Hardware Buttons):
Washing can also be performed without using the software. Washing parameters
can be stored on the injector by clicking Save as Default on the Wash tab of the
Injector Maintenance dialog box (in the Settings menu, click Injectors... and the
Injector Maintenance dialog box appears) of the i-control software. Press the
Prime/Wash button on the injector box to start the washing sequence using the
default parameters. (see Figure 43: Injector-box with injector in 'service position',
page 110). The injector must be connected the instrument and the instrument
must be switched on. Start the wash procedure by pressing and holding the
Prime/Wash button for more than 3 seconds.
THE INJECTOR CARRIER MUST BE IN THE SERVICE POSITION
DO NOT PERFORM WASHING WHEN THE INJECTOR IS IN THE
BE SURE TO RUN A FINAL WASH PROCEDURE WITH DISTILLED
WATER AND EMPTY THE INJECTOR SYSTEM. FOR GOOD CARE AND
LIFETIME FILL THE INJECTOR SYSTEM WITH LIQUID (WATER)
BEFORE TURNING OFF THE INSTRUMENT.
PLEASE SEE THE CORRESPONDING REAGENT KIT FOR ADVICE ON
HOW TO REMOVE THE SUBSTRATE COMPLETELY FROM THE
TAKE GOOD CARE OF THE INJECTORS, BECAUSE IF THEY ARE
DAMAGED THE ACCURACY OF DISPENSING MAY BE AFFECTED.
THIS CAN RESULT IN DAMAGE TO THE INSTRUMENT.
Injector needles can be replaced by exchanging the injector carrier
together with the corresponding tubing.
THE BUTTON(S) ON THE INJECTOR BOX INCLUDE TWO
PRESS THE BUTTON FOR LESS THAN 3 SECONDS TO
START PRIME.
PRESS THE BUTTON FOR MORE THAN 3 SECONDS TO
START WASH.
THE PARAMETERS HAVE TO BE SET IN THE i-control SOFTWARE.
IFU for Infinite 200 PRO No. 30052730
5. Operating the Instrument
CAUTION
FOR THE ACTION 'WASH'.
INSTRUMENT!
IMPORTANT
IMPORTANT
TUBING SYSTEM.
IMPORTANT
Note:
IMPORTANT
FUNCTIONS:
Rev. No. 1.6
115