Absorbance Baseline Flatness (Infinite F200 Pro) - tecan Infinite 200 PRO Instructions For Use Manual

Dna-expert reader
Table of Contents

Advertisement

7. Quality Control
7.3.9

Absorbance Baseline Flatness (Infinite F200 PRO)

Measurement Parameters:
Plate Layout:
Material/Reagents:
Calculation of Baseline Flatness:
172
Perform the following measurements to determine the baseline-flatness with the
available filter:
Parameter
Reading Mode
Measurement Wavelength
Number of flashes
Settle Time
Plate Type
Part of the Plate
Kinetic Cycles
No plate is necessary for measurement – the plate carrier has to be empty for this
measurement.
No material or reagents necessary for this test.
Calculate the standard deviation over 20 cycles for each wavelength.
IFU for Infinite 200 PRO
Setting
Absorbance
340 nm
405 nm
492 nm
590 nm
620 nm
700 nm
25
0
GRE96ft
A1
20, Minimal Interval Time
No. 30052730