Glow Type Luminescence; Atp Glow Luminescence - tecan infinite 200 Instructions For Use Manual

Table of Contents

Advertisement

7.6 Glow Type Luminescence

STOP
7.6.1

ATP Glow Luminescence

STOP
2008-07
Switch on the instrument at least 15 minutes before starting a
luminescence measurement. Some components need to warm up to
guarantee stable conditions for the measurement.
Luminescence Detection uses the photon counting technique.
Parameters
Wavelength Range
Lin. Dynamic Range
Integration Time/well
Cross Talk % (black plate)
Measurement range
Attenuation of Light
Plate Type (number of wells)
Total Dispensed Volume[µl]
Integration Time/well [ms]
ATP Detection Limit (Standard PMT)
ATP Detection Limit (Basic PMT)
Specifications are only valid when the instrument is placed in a
location with illumination < 500 lux.
Instructions for Use for infinite® 200 No. 30017581 Rev. No. 1.4
7. Instrument Features
Caution
Characteristics
380 – 600 nm
6 orders of magnitude
7 orders of magnitude (extended dynamic range)
1 – 20000 ms
< 0.01 %
> 6 orders of magnitude
10, 1 (no attenuation)
96
200
1000
3 fmol/well
20 fmol/well
Caution
121

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the infinite 200 and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF