6.5

Absorbance

2016-04
Parameters
Wavelength Range
Measurement
Range
The following specifications are valid for the wavelength range from
300 – 700 nm for Infinite 200 PRO series.
Plate type (number of wells)
Accuracy 0 – 2 OD
Accuracy 2 – 3 OD
Baseline Flatness
Infinite M200 PRO
Wavelength Accuracy
Infinite F200 PRO
Wavelength Accuracy
The specifications are valid for measurements performed with 25 flashes (reads)
per well.
IFU for Infinite 200 PRO No. 30052730
Instrument Type
Infinite F200 PRO
Infinite M200 PRO
Both
96
< ± (1 % + 10 mOD)
< ± 2.5 %
± 10 mOD (1 sigma)
≤ ± 1.5 nm
≤ ± 0.8 nm
Dependent on filters used
Rev. No. 1.6
6. Instrument Features
Characteristics
230 – 1000 nm
230 – 1000 nm
no filter necessary, selection
in 1 nm steps possible
0 – 4 OD
λ
> 315 nm;
λ
≤ 315 nm