Fluorescence Polarization (Infinite F200 PRO only)

Precision
Measurement Parameters:
2016-04
For the (Infinite F200 PRO) with the option 'Fluorescence Polarization' the
following tests may be performed to prove the specifications:
Precision/Repeatability
These test instructions are valid for the (Infinite F200 PRO):
Standard version
Spectrally enhanced version
Perform the following measurement to determine the precision:
Parameter
Reading Mode
Ex Wavelength
Em Wavelength
Number of flashes
Integration Time
Lag Time
Time between Move and Read
Gain
Plate Type
Reference from/to
Reference blank from/to
Reference value
Measurement blank from/to
IFU for Infinite 200 PRO No. 30052730
7. Quality Control
Setting
Fluorescence Polarization
485 (20) nm
535 (25) nm
25
40
0
0
Optimal
GRE96fb
A1 – D1
A2 – D2
20 mP
same as reference blank