Backflush Procedure (i-control):
Adjust parameters on the Backflush tab of the Injector Maintenance dialog box in
the Settings menu.
Start the reagent backflush procedure by clicking Start backflush.
11. Select one of the injectors Injector A or Injector B or Injector A and B (only
'primed' injectors are available for 'backflush').
12. Select the Piston Strokes (1 – 60; 1 stroke equals 1 ml)
13. Select the Backflush Speed (100 - 300 µl/sec).
14. Select the Refill Speed (100 – 300 µl/sec.) or select Refill Speed equal to
Backflush Speed.
15. Click Start backflush to start the reagent backflush procedure.
16. Click Close to exit the dialog box.
THE INJECTOR CARRIER MUST BE IN THE SERVICE POSITION
FOR THE ACTION 'BACKFLUSH'.
DO NOT PERFORM BACKFLUSH WHEN THE INJECTOR IS IN THE
IFU for Infinite 200 PRO No. 30052730