6.7
Flash Type Luminescence
STOP
STOP
2016-04
SWITCH ON THE INSTRUMENT AT LEAST 15 MINUTES BEFORE
STARTING A LUMINESCENCE MEASUREMENT. SOME COMPONENTS
NEED TO WARM UP TO GUARANTEE STABLE CONDITIONS FOR THE
Luminescence Detection uses the photon counting technique.
Parameters
Wavelength Range
Measurement Range
Integration Time/well
Cross talk % (black plate)
Attenuation of Light
ATP Detection Limit
SPECIFICATIONS ARE ONLY VALID WHEN THE INSTRUMENT IS
PLACED IN A LOCATION WITH ILLUMINATION < 500 LUX.
IFU for Infinite 200 PRO No. 30052730
6. Instrument Features
CAUTION
MEASUREMENT.
Characteristics
380 – 600 nm
> 6 orders of magnitude
8 orders of magnitude
(extended dynamic range)
100 – 20000 ms
< 0.01 %
100 (OD2 attenuation filter),
1 (no attenuation)
< 80 amol/well
CAUTION
Rev. No. 1.6
143