7. Quality Control
Precision
Measurement Parameters:
Plate Layout:
<>
1
2
A
B
C
D
E
F
G
H
Material/Reagents:
Calculation of Precision:
160
Perform the following measurement to determine the precision/reproducibility:
Parameter
Reading Mode
Ex Wavelength
Em Wavelength
Number of flashes
Integration Time
Settle Time
Gain
Plate Type
Part of the Plate
Kinetic
Interval Time
Pipette 200 µl of 1 nM Fluorescein or the blank solution (0.01 M NaOH) into the
appropriate wells according to the plate layout:
3
4
5
1 nM Fluorescein (in 0.01 M NaOH) (Fluorescein sodium salt, Sigma F6377)
0.01 M NaOH (=Blank) (NaOH pellets, Merck article no. 6495 or Sigma S8045)
1 Greiner 96-well plate black
200 µl Pipette + tips
stdev
=
Pr
ecision
(
CV
%)
mean
Average RFU value of well A1 over the 20 kinetic
wellA1
stdev
Standard deviation of RFU values of Well A1 over the 20 cycles
wellA1
The result of the formula determines the Precision in % CV.