7.3.8
Absorbance Baseline Flatness (Infinite M200 PRO)
Measurement Parameters:
Plate Layout:
Material/Reagents:
Calculation of Baseline Flatness:
2016-04
Perform the following measurement to determine the baseline-flatness:
Parameter
Reading Mode
Measurement Wavelength from/to
Number of flashes
Settle Time
Plate Type
Part of the Plate
No plate is necessary for measurement – the plate carrier has to be empty for this
measurement.
No material or reagents necessary for this test.
Calculate the standard deviation.
IFU for Infinite 200 PRO No. 30052730
7. Quality Control
Setting
Absorbance
300 - 700 nm
25
0
GRE96ft
A1
Rev. No. 1.6
171