7.3
Specifications - Test Instructions
7.3.1
Fluorescence Top
Sensitivity
2016-04
For the Infinite 200 PRO with the option 'Fluorescence Top', the following tests
can be performed to prove the specifications:
•
Sensitivity
•
Uniformity
•
Precision
These test instructions are valid for the Infinite 200 PRO:
•
Infinite F200 PRO
•
Infinite M200 PRO
•
Standard version
•
Spectrally enhanced version
Perform the following measurement to determine the detection limit for
Fluorescein:
Measurement Parameters:
Parameter
Reading Mode
Ex Wavelength
Em Wavelength
Number of flashes
Integration Time
Settle Time
Gain
Plate Type
IFU for Infinite 200 PRO No. 30052730
7. Quality Control
Setting
Fluorescence Top
Infinite F200 PRO: 485 (20) nm
Infinite M200 PRO: 485 nm
Infinite F200 PRO: 535 (25) nm
Infinite M200 PRO: 535 nm
25
40
0
Optimal
GRE96fb
Rev. No. 1.6
157
Need help?
Do you have a question about the Infinite 200 PRO and is the answer not in the manual?
Questions and answers