Error #
Error Text
39
Injector A Communication
Timeout
40
Injector B Communication
Timeout
41
Prime Wash Error
42
Instrument is locked
43
Prepare: [channel]:
Wavelength:[lambda]
Gain:[g], Counts: [cts]
44
Steploss Error
45
Sync Scan: Number of EX-
Steps does not match EM-
Steps
46
Handshake timeout at
module
47
Motor Timeout
48
[Value] is not in defined a
Range
49
Sensor is broken
2016-04
Description
Communication Error on Injector Interface. Please report
this error to your local customer support office.
Communication Error on Injector Interface. Please report
this error to your local customer support office.
Injectors still priming or washing. Please wait until prime or
wash process is finished.
Instrument is locked after a serious hardware problem. For
unlocking a reboot is necessary. Please report this error to
your local customer support office.
Unspecific Hardware failure. Please report this error to your
local Tecan customer support office.
Actuator failure. Please report this error to your local Tecan
customer support office.
Unspecific error in the Instrument - Computer
communication protocol. Please report this error to your
local Tecan customer support office.
Unspecific Hardware Failure. Please report this error to
your local Tecan customer support office.
Unspecific Hardware Failure. Please report this error to
your local Tecan customer support office.
Unspecific Hardware Failure. Please report this error to
your local Tecan customer support office.
Sensor Failure. Please report this error to your local Tecan
customer support office.
IFU for Infinite 200 PRO No. 30052730
9. Troubleshooting
Rev. No. 1.6
187