4. Gas Control Module (Enhanced)
4.2
Gas Control Module Features
4.2.1
Gas Control Module Configurations
'CO2' and 'O2' Configuration
54
The gas control module enhanced (GCM
variety of cell-based applications for the Infinite 200 PRO multimode reader.
Two integrated gas inlets allow the external control of CO
maintain stable culture conditions and improve cell growth. Carbon dioxide
concentration is regulated by an inflow of CO
achieved by supplying N
The GCM makes the Infinite 200 PRO ideally suited for in vitro studies of
eukaryotic cell lines or extends the usage of the instrument for microplate
investigations using anaerobe or facultative anaerobic bacteria.
Settings:
The O
concentration can be set in the range from 0.1 – 21 vol. %.
2
The O
control range is 0.1 – 21 vol. %.
2
The CO
concentration can be set in the range from 0 – 10 vol. %.
2
The CO
control range is 0 – 10vol. %.
2
The Accuracy of measured O
The Accuracy of measured CO
The GCM is available in two configurations:
•
'CO2 and O2' configuration: CO
regulated inside the reader chamber.
•
'CO2' configuration: CO
chamber
There are four gas control modes available for the GCM enhanced with
'CO2 and O2' configuration:
•
Carbon Dioxide Mode: In CO2 Mode, CO
instrument (plate carrier compartment/reader chamber).
•
Oxygen Mode: In O2 Mode, N2 gas is introduced to reduce the oxygen gas
concentration inside the instrument to below ambient air levels.
•
DUAL Mode: In DUAL Mode, the flow of CO
automatically by opening or closing the valve while monitoring both
concentrations (vol. %) inside the instrument.
•
Manual Mode: In Manual Mode, the flow of CO
by opening or closing the valve manually while monitoring both
concentrations (vol. %) inside the instrument.
IFU for Infinite 200 PRO
TM
) offers a comprehensive solution for a
gas, whereas oxygen reduction is
2
gas.
2
concentration is +/- 0.9 vol. %.
2
concentration is +/- 1 vol. %.
2
and/or O
2
concentration can be regulated inside the reader
2
2
No. 30052730
Rev. No. 1.6
and O
to help
2
2
concentrations can be
2
gas is introduced into the
and N
gas is controlled
2
2
and/or N
gas is controlled
2
2
2016-04
Need help?
Do you have a question about the Infinite 200 PRO and is the answer not in the manual?
Questions and answers