On The Fly" Measurements - tecan Infinite 200 PRO Instructions For Use Manual

Dna-expert reader
Table of Contents

Advertisement

6.10

"On the Fly" Measurements

2016-04
AlphaScreen/AlphaLISA measurements are only possible as endpoint
measurements in white or light gray microplates and cannot be
performed in combination with the injector system and the heating
Parameters
Wavelength Range:
Standard Filters
AlphaScreen:
Standard Filters
AlphaLISA:
Gain Setting
Manual
Calculated from
Well
Parameters
Excitation time:
Integration time:
"On the Fly" measurements are the fastest measurements possible using the
Infinite 200 PRO. These measurements are performed using only one flash
(number of flashes).
96-well plates (FI, TRF, Absorbance) Measurement time: < 20 s
384-well plates (FI, TRF, Absorbance) Measurement time: < 30 s
(Plate-in/out movement not included).
IFU for Infinite 200 PRO No. 30052730
Note
system.
Characteristics
380 - 700 nm
Excitation: 680 (30) nm
Emission: 570 (100) nm
Excitation 680 (30) nm
Emission: 615 (20) nm
Values
Measurement Range
1 - 255
0 - 60,000 RFU
automatic
0 - 60,000 RFU
Characteristics
100 - 2000 ms (in 10 ms steps)
100 - 2000 ms (in 10 ms steps)
Rev. No. 1.6
6. Instrument Features
145

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Infinite 200 PRO and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Table of Contents

Save PDF