5. Operating the Instrument
5.2
General Operating Features
5.2.1
Instrument Start Up
Instrument Power On
Connect to Instrument
5.3
General Options
76
The Infinite 200 PRO has some general behavior and options, which are
independent from a particularly selected measurement technique.
Before the instrument is switched ON, check if the USB interface cable is
connected.
When switching ON the instrument no initialization steps are performed.
When the software connects to the instrument, communication is established
between the instrument and the user interface.
The following steps are performed:
1.
OS filter wheels are initialized (M200 only)
2.
Luminescence filter wheel is initialized
3.
Z-transport of luminescence optics is initialized
4.
Plate transport is initialized
(The plate transport is not moved out automatically)
5.
The current versions of firmware and software are displayed.
6.
The instrument is ready for use.
The following options may be taken independently from the particular
measurement technique.
To keep temperature on a constant level and provide uniformity
across the plate, the plate must be placed in "incubation position".
When the "heating" function is used during shaking,
the temperature may vary slightly.
IFU for Infinite 200 PRO
Note
No. 30052730
Rev. No. 1.6
2016-04