4. Gas Control Module (Enhanced)
4.3.2

O2 Mode

58
In O2 Mode, it is possible to regulate O
value and starting the gas control module. (Nitrogen gas is introduced to reduce
the oxygen gas concentration inside the instrument to below ambient air levels).
1.
Select the mode by pressing the 'I/O' key, the following display appears:
2.
Change target value for 'O2' (using the plus and minus keys)
The editable target value blinks (highlighted in gray above)
3.
Press the 'I/O' key to Start/Stop gas regulation. The 'actual' value shows
the current gas concentration inside the reader. (LED blinks when gas
regulation is running - LED is stable gas regulation is stopped).
4.
Press the 'Mode' key for 1.5 seconds to return to the 'O2 Mode' display
(main menu).
IFU for Infinite 200 PRO
gas concentration by setting the target
2
No. 30052730
Rev. No. 1.6
2016-04