Optimizing Luminescence Measurements; Integration Time; Light Level Attenuation - tecan Infinite 200 PRO Instructions For Use Manual

Dna-expert reader
Table of Contents

Advertisement

5.9

Optimizing Luminescence Measurements

STOP
5.9.1

Integration Time

5.9.2

Light Level Attenuation

2016-04
SWITCH ON THE INSTRUMENT AT LEAST 15 MINUTES BEFORE
STARTING A LUMINESCENCE MEASUREMENT. SOME COMPONENTS
NEED TO WARM UP TO GUARANTEE STABLE CONDITIONS FOR THE
At very low light levels, a PMT does not yield a continuous output current, which
is necessary for a reliable analog to digital conversion. Instead, it produces a
sequence of pulses the average rate of which can be measured using a counter.
The advantage of the photon counting technique at such low light levels is that
pulse height selection criteria allow electronic noise to be discriminated.
At very low light levels the measured counts per second are proportional to the
light intensity. Increase of measurement time per well yields more accurate
values because of the irregular photon impact (photon statistics). The photonic
noise (shot noise) cannot be reduced technically.
The signal to noise (S/N) ratio can be optimized by increasing the
integration time. Increasing the integration time by a factor of 10
results in an improvement of the S/N ratio by app. a factor of 3.
When using photon counting detection, optical attenuation of higher
luminescence light levels (>10,000,000 counts per second) is necessary. In such
a case, too many photons enter the luminescence detector at one time and
cannot be distinguished as distinct exit pulses. Count rates would even fall behind
values at lower light levels.
Therefore, values >10,000,000 counts per second (without attenuation) are
displayed as "INVALID" in the results sheet.
The optical system of the Infinite 200 PRO can attenuate light levels by a fixed
factor of either 1 (none) or 100 (2 OD). Correspondingly, the usable
measurement range will be shifted to higher light levels (<1,000,000,000 counts
per second).
IFU for Infinite 200 PRO No. 30052730
5. Operating the Instrument
CAUTION
MEASUREMENT.
Note
Rev. No. 1.6
107