tecan Infinite 200 PRO Instructions For Use Manual page 43

Dna-expert reader
Table of Contents

Advertisement

Absorbance Optics Cuvette
Absorbance Detection Cuvette
Cuvette types
Cuvette Type
Width x Depth
Standard
12.5 x 12.5 mm
cuvettes
Semi-macro
12.5 x 12.5 mm
cuvettes
Micro
12.5 x 12.5 mm
cuvettes
Ultra-micro
12.5 x 12.5 mm
cuvettes
Hellma GmbH & Co. KG, Germany; www.hellma-worldwide.com
2016-04
The cuvette port is an option of the Infinite M200 PRO only. This option
is not available for the Infinite F200 PRO. With the Infinite F200 PRO,
cuvettes may be measured using a Tecan Cuvette Adapter placed on
A fiber bundle guides the light from the excitation monochromator system to the
absorbance cuvette optics.
This optics consists of a pair of lenses focusing the light beam into the cuvette.
At the focal point, the spot diameter of the absorbance light beam is 1.9 mm.
A silicon photodiode is used for the measurement of the transmitted light. It is
sensitive to a wide range of wavelengths. The photodiode is well suited for the
light levels being encountered with absorbance measurements below 4 OD.
Measurement values above 4 OD are marked as 'OVER' in the result sheet.
The cuvette port is compatible with the following cuvettes:
(including lid)
Cuvettes with a measurement window < 2 mm (diameter) cannot be used.
IFU for Infinite 200 PRO No. 30052730
Note
the plate transport.
Maximum
Filling
Height
Volume
55 mm
2 ml
55 mm
1 ml
55 mm
0.5 ml
55 mm
100 µl
Rev. No. 1.6
2. General Description
Example
Hellma 110 QS,
10 mm
Hellma 108-QS,
10 mm
Hellma 104.002 QS,
10 mm
Hellma 105.202,
10 mm
43