6. Instrument Features
6.6
Glow Type Luminescence
STOP
6.6.1
ATP Glow Luminescence
STOP
142
SWITCH ON THE INSTRUMENT AT LEAST 15 MINUTES BEFORE
STARTING A LUMINESCENCE MEASUREMENT. SOME COMPONENTS
NEED TO WARM UP TO GUARANTEE STABLE CONDITIONS FOR THE
Luminescence Detection uses the photon counting technique.
Parameters
Wavelength Range
Linear Dynamic Range
Integration Time/well
Cross Talk % (black plate)
Measurement range
Attenuation of Light
Plate Type (number of wells)
Total Dispensed Volume[µl]
Integration Time/well [ms]
ATP Detection Limit
SPECIFICATIONS ARE ONLY VALID WHEN THE INSTRUMENT IS
PLACED IN A LOCATION WITH ILLUMINATION < 500 LUX.
IFU for Infinite 200 PRO
CAUTION
MEASUREMENT.
Characteristics
380 – 600 nm
6 orders of magnitude
100 – 20000 ms
< 0.01 %
> 6 orders of magnitude
8 orders of magnitude
(extended dynamic range)
100 (OD2 attenuation filter),
1 (no attenuation)
96
200
1000
3 fmol/well
CAUTION
No. 30052730
Rev. No. 1.6
2016-04