G-Factor Settings; Measurement With An Uncalibrated G-Factor - tecan Infinite 200 PRO Instructions For Use Manual

Dna-expert reader
Table of Contents

Advertisement

5.6.3

G-Factor Settings

STOP
5.6.4

Measurement with an Uncalibrated G-Factor

2016-04
The given equation for calculation of fluorescence polarization assumes that the
sensitivity of the detection system is equivalent for parallel and perpendicular
polarized light. This is generally not the case and either the parallel or
perpendicular intensity must be corrected by so called 'G-Factor'. The G-factor
compensates for differences in optical components between parallel and
perpendicular measurement.
The G-Factor is the correction factor that can be determined for the wavelength of
the fluorophore by measuring a sample with a known polarization value. A valid
calibration of the instrument resulting in a G-factor is an important requirement for
each fluorescence polarization measurement.
MAKE SURE THAT THE FILTER SLIDE CONTAINS POLARIZERS
TOGETHER WITH THE FILTERS DEFINED FOR FLUORESCENCE
POLARIZATION. MEASUREMENTS WITHOUT THE POLARIZERES WILL
RESULT IN A FALSE G-FACTOR AND FALSE MEASUREMENT DATA.
If no calibrated G-factor is available, the default value of 1 will be displayed and
marked as 'Uncalibrated G-Factor'. In order to enable the measurement, confirm
this value or select a new one by either clicking the up and down arrows or
entering a value in the G-Factor field.
For the G-Factor calibration, see 5.6.5.
IFU for Infinite 200 PRO No. 30052730
5. Operating the Instrument
CAUTION
Rev. No. 1.6
97

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Infinite 200 PRO and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents

Save PDF