The injector carrier, which includes the injector needles, can be easily removed
from the instrument for priming or washing the system and for optimizing the
injection speed.
Figure 6: Injector carrier
When using the injector during a measurement or for just dispensing a plate the
injector carrier must be inserted correctly into the instrument. Remove the injector
dummy and insert the carrier into the injector port. Press the carrier softly into the
injector port until you hear a clicking noise.
The instrument contains an injector sensor that checks that the position of the
injector carrier for the actions 'inject' and 'dispense' is correct.
If the injector carrier is not inserted correctly, the injector sensor does not
recognize the inserted carrier and neither dispensing nor injection is possible.
On the other hand, actions like washing and priming are enabled although the
injector carrier is inserted; therefore, always make sure that the injector carrier is
in the service position for washing and priming.
THE INJECTOR CARRIER MUST BE IN THE SERVICE POSITION
PRIME AND WASH MUST NOT BE PERFORMED
WHEN THE INJECTOR IS IN THE INSTRUMENT!
IFU for Infinite 200 PRO
CAUTION
FOR WASHING UND PRIMING.