Measurement with a Calibrated G-Factor

2016-04
The calculated G-Factor is automatically stored on the computer's hard drive.
Each G-Factor entry corresponds to the filter pair selection as well as the filter
slide description. There is always only one G-Factor available for the respective
filter pair combination and filter slide description, unless the same filter pair has
been used with the different filter slides and thus stored with the different filter
slide descriptions.
The filter slide description is part of the G-Factor key value. Avoid
using the same filter slide description for different filter slides as this
will affect the correct G-Factor recognition.
Once calibrated, the G-factor is shown and can be used immediately if
it matches the Ex/Em wavelength pair and the filter slide description.
A calibrated G-factor will be displayed automatically or can be loaded by clicking
the '>>' button only if it matches the selected fluorescence polarization filter pair
and the filter slide description.