8. Cleaning and Maintenance
8.4

Instrument Disinfection

8.4.1

Disinfection Solutions

180
All parts of the instrument that come into contact with the patient samples,
positive control samples or hazardous material must be treated as potentially
infectious areas.
THE DISINFECTION PROCEDURE SHOULD BE PERFORMED
ACCORDING TO NATIONAL, REGIONAL, AND LOCAL
ALL PARTS OF THE INSTRUMENT THAT COME INTO CONTACT
WITH POTENTIALLY INFECTIOUS MATERIAL MUST BE TREATED
AS POTENTIALLY INFECTIOUS AREAS.
IT IS ADVISABLE TO ADHERE TO APPLICABLE SAFETY
PRECAUTIONS, (INCLUDING THE WEARING OF POWDER-FREE
GLOVES, SAFETY GLASSES AND PROTECTIVE CLOTHING) TO
AVOID POTENTIAL INFECTIOUS DISEASE CONTAMINATION WHEN
PERFORMING THE DISINFECTION PROCEDURE.
Before the instrument is returned to the distributor for servicing, it must be
disinfected and a disinfection certificate completed. If a disinfection certificate is
not supplied, the instrument may not be accepted by the servicing center or it
may be held by the customs authorities.
The instrument should be disinfected using the following solution:
B30 (Orochemie, Max-Planck-Str. 27; D-70806 Kornwestheim)
IFU for Infinite 200 PRO
WARNING
REGULATIONS.
WARNING
No. 30052730
Rev. No. 1.6
2016-04