Cuvette
Absorbance strip
(cuvette)
Part of Plate
(not shown)
Absorbance strip
(microplate)
2016-04
Example 2:
Combination of microplate and cuvette measurement:
For some applications it might be necessary to compare data measured on a
microplate with cuvette data. The following example shows how to set up this
measurement in general:
Necessary for cuvette measurements.
Up to 4 absorbance fixed wavelength strips are allowed. Reference
wavelength is only selectable when using one absorbance fixed
wavelength strip. 'Apply blanking' is disabled when a reference
wavelength is selected.
Select the appropriate measurement parameters (wavelength,
number of flashes and settle time)
Necessary for microplate measurements. Select an appropriate
Plate
plate type for the measurement.
Optional. Use the 'part of plate' strip if only a part of the plate shall
be measured.
Up to 10 absorbance fixed wavelength strips are allowed.
Reference wavelength is only allowed on the first absorbance strip.
Reference wavelength is disabled on absorbance strips 2 to 10.
Select the appropriate measurement parameters (wavelength,
number of flashes and settle time) for your application.
IFU for Infinite 200 PRO No. 30052730
5. Operating the Instrument
Rev. No. 1.6
127