Replacing Fuses

Introduction
Procedure
6-4 Maintenance
All instruments have factory-installed fuses. If you need to change the fuses, use the
procedure below.
FIRE HAZARD.
replace fuses only with fuses of the type and rating specified for the instrument
The fuse compartment is located at the instrument rear, as shown below.
GR1990
Items Needed
Two fuses of the same type you are removing
Fine flat-tip screwdriver
To replace the fuses:
Step
Action
1
Turn off the 2720 Thermal Cycler and disconnect the power cord from the
instrument rear.
which could cause physical injury or death, can result from working on an
instrument when the high-voltage power supply is operating. To avoid
electrical shock, disconnect the power cord to the instrument before
opening the fuse compartment.
Wait 30 seconds before any further work to let any electrical charges dissipate.