Calibrate Gas Thermistor; Water Bath; Resistor; Calibrate Oven Thermistor - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Calibrate Oven Thermistor

Thermo Fisher Scientific

Water Bath

The Calibrate Gas Thermistor Water Bath screen is used to view and set
the permeation oven gas temperature to a known temperature value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
Thermistor > Water Bath.

Resistor

The Calibrate Gas Thermistor Resistor screen is used to view and set the
permeation oven gas thermistor resistance to a known resistor value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
Thermistor > Known Resistor.
The Calibrate Oven Thermistor screen is used to view and set the
permeation oven thermistor resistance to a known resistor value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Oven
Thermistor.

CALIBRATE GAS THERMISTOR:

>WATER BATH
KNOWN RESISTOR
RANGE
AVG
DIAGS ALARM
CAL GAS THERM (BATH):
CURRENTLY:
SET TO:

MOVE CURSOR

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
CAL GAS THERM (RESISTOR):
CURRENTLY:
SET TO:

MOVE CURSOR

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
o
45.80
C
o
45.00
C
SAVE
3850 Ohms
04000
Ohms
SAVE
Model 42i Instruction Manual
Operation
Service Menu
3-87