Calibrate Oven Thermistor
Thermo Fisher Scientific
Water Bath
The Calibrate Gas Thermistor Water Bath screen is used to view and set
the permeation oven gas temperature to a known temperature value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
●
Thermistor > Water Bath.
Resistor
The Calibrate Gas Thermistor Resistor screen is used to view and set the
permeation oven gas thermistor resistance to a known resistor value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Gas
●
Thermistor > Known Resistor.
The Calibrate Oven Thermistor screen is used to view and set the
permeation oven thermistor resistance to a known resistor value.
In the Main Menu, choose Service > Perm Oven Settings > Cal Oven
●
Thermistor.
CALIBRATE GAS THERMISTOR:
>WATER BATH
KNOWN RESISTOR
RANGE
AVG
DIAGS ALARM
CAL GAS THERM (BATH):
CURRENTLY:
SET TO:
MOVE CURSOR
CHANGE VALUE
RANGE
AVG
DIAGS ALARM
CAL GAS THERM (RESISTOR):
CURRENTLY:
SET TO:
MOVE CURSOR
CHANGE VALUE
RANGE
AVG
DIAGS ALARM
o
45.80
C
o
45.00
C
SAVE
3850 Ohms
04000
Ohms
SAVE
Model 42i Instruction Manual
Operation
Service Menu
3-87